DATELINE 27 MAY 2021: Wow! Some companies made it just across the line in time and I’m sure some didn’t. Did you? Not everyone agress with my perspective. Let’s get on with the story.

What does that all mean?

The MDD and AIMDD are dead (Dead on Arrival) and the MDR Date of Application (DoA) has just flown by us, yesterday. So, we are past the transition period, per say, for the MDR and now we’re into a grace (grey zone) for the legacy MDD/AIMDD products that can’t have any significant changes under their existing MDD/AIMDD certificates otherwise they need to go “full up” into MDR-land.

IEC 60601-1-2:2020 (Ed. 4.1) Hidden Surprise!

IEC 60601-1-2:2020 (Ed. 4.1) Hidden Surprise!

IEC 60601-1-2:2014, 4th edition, references CISPR 11:2010 (a.k.a. EN 55011), had no radiated rf emissions requirements for group 2 devices (intentional radiators) between 150kHz and 30 MHz. This is the basis for all group 2 medical devices on the market today.
IEC 60601-1-2:2020, edition 4.1, got fast tracked by the FDA since it contains new immunity tests (table 11 – RFID requirements based on AIM 735131 standard values) that they want manufacturers to do immediately. In that 2020 standard, 60601-1-2 references CISPR 11:2015 + A1:2016 + A2:2019 version that inherits a new group 2 only RF radiated emissions test with limits that apply for 150kHz – 30 MHz.